Measuring - Percentile Analysis
Vorheriges Thema  Nächstes Thema 

Percentile Analysis:


Click <perc> to enable the percentile analysis.

Click <go> or press the Enter key to start the actual measurement.


Click <go> again or press the ESC key to interrupt or cancel the measurement.


Figure 62: Percentile analysis option window

Selection window options (Figure 62):

  • Signal type
  • Input level
  • Lead time: time before the measurement starts
  • Measuring time: relevant only in the setting for one-time
  • Click <Turn on signal 2> to enable Signal 2.
    • Present a second signal to the loudspeaker (or several speakers for real ear).
    • When Signal 2 is enabled, the signal type and level may be adjusted.


The measurement will appear on the screen after it is completed. Selected curves can be edited by clicking: <insert><hide>, or <delete>.

For further information, please refer to the section: "Display and administrate curves."


Various display modes in the percentile analysis:


Settings for the display modes are located in the Presentation tab.


Figure 63a: 3 values model

Figure 63b: 3 values model


Figure 64a: 4 values model


Figure 64b: 4 values model


Figure 65a: 5 values model


Figure 65a: 5 values model


Figure 66a: Display of diagram targets, dynamic compression rates (CR), and the target curve transformation


Figure 66b: Display of the diagram targets (red), the dynamic CR (numerical values at the percentile bars) and the target curve transformation (black line).


Figure 67a: Display at lines


Figure 67b: Display at lines


Percentile analysis measuring options:


Figure 68: Percentile analysis window


Table refers to Figure 68.

Setting

Description

Signal Type

  • Select the signal type.
  • Signal types include:
    • Speech
    • Natural signals
    • Technical signals (e.g. CHIRP)


Input Level

  • Select the input level.
  • If the desired level is not available on the drop-down menu, manually enter the value in the field.


Modulation Frequency and Modulation Depth

  • An amplitude modulation may be selected for the signal types: CHIRP, white noise, pink noise, speech simulated noise, or pure tone.
  • Can be used to determine dynamic compression rates in hearing aids.


Lead Time

  • Time between when the measurement is enabled and the actual start of the measurement.
  • The actual start of the measurement is delayed.
  • Lead Time only applies when performing a single measurement and when the hearing aid has long attack times.


Measuring Time

  • Only applies when performing a single measurement.

Enable Signal 2

  • Add a second signal to the actual measuring signal.
  • May be used to analyze noise reduction algorithms in hearing aids or to test directional microphones (only available in real ear).


Signal 2 Signal Type

  • Identical to “Signal Type” above.


Signal 2 Level

  • Identical to “Input Level” above.


Signal 2 to Loudspeaker

  • Only enabled in real ear.
  • It is possible to present a second signal from a different direction, provided that the corresponding loudspeaker is activated and connected.


Start

  • Starts the actual measurement.


Abort

  • Measurement does not start


Calibration of percentile analysis:


To determine if calibration is necessary, observe the total level of the reference micrphone during the calibration meaurement (Figure 69). This value should match the level adjusted by the user. If this is not correct, calibration should be performed of the signals that are being used. Otherwise, it will take 15 minutes to calibrate all available signals.


Figure 69: Calibrating the percentile analysis



Selection of the signals for calibration:


The ACAM System estimates the amount of time necessary for calibration and displays an approximate value. Figure 70 shows an estimated time of 1 minute.


In this window, you can click on the box to select and deselect signals to be calibrated. Click the <start calibration> button to begin the calibration.


Figure 71: Selection of signals for calibration


Own Files:


Figure 72: Own Files window


Adjustments in Setup:


The measuring accuracy/measuring speed of the percentile analysis can be adjusted in the setup window of the Measure tab.


Select from the following resolutions:

Resolution

Number of measuring points (Defaults)

Very good

500 Measurements

Good

200 Measurements

Medium

100 Measurements

Fast

40 Measurements

Very fast

20 Measurements


Note: